Recommended Flow Rates for Gas Detection Instruments*

When calibrating or performing a bump test, it is important to use a regulator with flow rate that falls within the manufacturer's recommendations. Pumped instruments may generally be used with a demand flow regulator (DFR), which is the most accurate and efficient way to apply gas to the sensors.

Please consult your gas detector manual for more information on your specific gas monitor and bump test and calibration procedures.

Draeger PAC 5000, 7000 0.5 LPM
X-AM 2500 0.5 LPM
Honeywell Analytics (BW) GasAlertMicroClip 0.5 LPM
GasAlertMax XT II 0.5 LPM, DFR
GasAlert Micro 5 0.5 LPM
MSA Altair 4X 0.25 LPM
Altair 5X DFR
RAE Systems ToxiRAE Pro 0.5 to 1.0 LPM
QRAE 3 0.5 to 1.0, DFR
MicroRAE 0.5 to 1.0 LPM
MultiRAE 0.5 to 1.0 LPM, DFR
RKI Instruments GX-2009 0.5 LPM
GX-6000 DFR
  • DFR = Demand Flow Regulator
  • LPM = Liters Per Minute

* For reference purposes only. Consult gas detector manual for exact manufacturer recommendations.