Nitrogen Oxides Detection Tubes (5 to 625 ppm)
Gastec Detection Tube No. 11S measures Nitrogen Oxides with a Range of 5 to 625 ppm.
- Measuring Range:
- 5 to 10 ppm
- (10) to 250 ppm
- 250 to 625 ppm
- Number of Pump Strokes: 2, 1, 1/2
- Correction Factor: 1/2, 1, 2.5
- Sampling Time: 45 seconds per pump stroke
- Detecting Limit: 2 ppm (2 pump strokes)
- Changes from White to Pale Green
- Corrections for temperature & humidity: Unnecessary
- Reaction Principle:
- NO + Cr6+ + H2SO4 ? NO2
- NO2 + (C6H5)2NH ? C6H5NHC6H4NO
- Relative Standard Deviation: 10% (for 10 to 80 ppm), 5% (for 80 to 250 ppm)
- 2 Year Shelf Life
- Store the tubes at cool and dark place.
Possible coexisting substances and their interferences:
| Substance |
Concentration |
Interference |
Changes color by itself to |
| Hydrogen chloride |
? 50 ppm |
Unclear demarcation |
Bluish purple (? 10 ppm) |
| Hydrogen sulfide |
? 1/1 |
+ |
No |
| Ozone |
? 80 ppm |
Unclear demarcation (Two layers) |
Pale brown |
| Sulfur dioxide |
? 1/1 |
+ |
No |
| Methanol |
? 400 ppm |
- |
No |
| Nitric oxide is oxidized to form nitrogen dioxide. If organic solvent of high concentration is coexisting, oxidising agent is deteriorated to produce minus error for Nitric oxide concentration. |
Nitrogen Oxides Detection Tubes (5 to 625 ppm)
There are no reviews yet.