Ethylene Oxide Detection Tubes 25/a (25 - 500 ppm)
For short-term measurements more than 160 Draeger-Tubes are available to measure the spot concentrations of specific gases. Measurement usually takes between 4 seconds and 15 minutes. The tubes are used either manually with the Draeger-Tube Accuro Pump or automatically with the Draeger X-act 5000.
A wide range of different gases and vapors can be detected and measured, such as:
- Determining and measuring concentration peaks
- Measuring personal exposure in the inhalation area
- Detecting leaks in pipelines, gas installations etc.
- Air analysis in sewers, shafts and confined spaces
Ethylene Oxide Detection Tubes 25/a (25 - 500 ppm)
Standard Measuring Range: |
25 to 500 ppm |
Number of Strokes (n): |
30 |
Time for Measurement: |
approx. 6 min |
Standard Deviation: |
�20 to 30 % |
Color Change: |
pale yellow ?> turquoise green |
Temperature: |
15 to 30� C |
Absolute Humidity: |
3 to 15 mg H2O / L |
Reaction Principle: |
Ethylene oxide + CrVI ??> CrIII + various oxidation products |
Cross Sensitivity: |
Alcohols, esters and aldehydes are also indicated, but with different sensitivities. It is impossible to differentiate between them. Propylene oxide is also indicated, but with a different sensitivity. Ethylene, ketones and toluene in the TLV range do not interfere. |
Ethylene Oxide Detection Tubes 25/a (25 - 500 ppm)