Toluene Detection Tubes 5/b (5 - 300 ppm)

There are no reviews yet. | Write a review
Item #: 8101661
MFG: Draeger
UOM: Box of 10
Min Order: 0
Order Qty: Must order in multiples of 1
$102.00
View additional buying options

Toluene Detection Tubes 5/b (5 - 300 ppm)

For short-term measurements more than 160 Draeger Tubes are available to measure the so called spot concentrations of specific gases.

A wide range of different gases and vapors can be detected and measured, such as:

  • Determining and measuring concentration peaks
  • Measuring personal exposure in the inhalation area
  • Detecting leaks in pipelines, gas installations etc.
  • Air analysis in sewers, shafts and confined spaces
Standard Measuring Range: 50 to 300 ppm / 5 to 80 ppm
Number of strokes n: 2 / 10
Time for Measurement: approx. 2 min / approx. 10 min
Standard Deviation: ± 10 to 15 %
Color Change: white ──> pale brown
Temperature: 2 to 40 °C
Absolute Humidity: max. 20 mg H2O / L
Reaction Principle: Toluene + I2O5 + H2SO4  ──> I2
Cross Sensitivity: 10 ppm phenol, 1,000 ppm acetone, 1,000 ppm ethanol and 300 ppm octane are not indicated, xylene (all isomers) and benzene are indicated with the same sensitivity. The discoloration in the presence of p-xylene is violet, and yellowish-green with benzene.