Perchloroethylene Detection Tubes 2/a (2 - 300 ppm)
Standard Measuring Range: |
20 to 300 ppm / 2 to 40 ppm |
Number of strokes n: |
1 / 5 |
Time for Measurement: |
approx. 30 sec / approx. 3 min |
Standard Deviation: |
� 15 to 20 % |
Color Change: |
slightly greenish ??> grey blue |
Temperature: |
15 to 30 �C |
Absolute Humidity: |
< 25 mg H2O / L |
Reaction Principle: |
a) CCl2=CCl2 + MnO4- ??> Cl2 b) Cl2 + Diphenylbenzidine ??> gray blue reaction product At higher concentrations there may be created a reddish area at the beginning of the indicating layer. |
Cross Sensitivity: |
Other chlorinated hydrocarbons, free halogens and hydrogen halides are also indicated. Petroleum hydrocarbon vapors result in a shortened reading once they exceed the following concentrations: 50 ppm with 5 strokes or 500 ppm with 1 strokes |