Cyclohexanone Detection Tubes (2 to 75 ppm)

There are no reviews yet. | Write a review
Item #: 154
MFG: GasTec
UOM: Box of 10
Min Order: 0
Order Qty: Must order in multiples of 1
$98.81
View additional buying options
Product Details

Cyclohexanone Detection Tubes (2 to 75 ppm)

Gastec Detection Tube No. 154 measures Cyclohexanone with a Range of 2 to 75 ppm.

  • Measuring Range: 2 to 30 ppm, 30 to 75 ppm
  • Number of Pump Strokes, 4, 2
  • Correction Factor: 1, 2.5
  • Sampling Time: 3 minutes per pump stroke
  • Detecting Limit: 0.2 ppm (4 pump strokes)
  • Changes from Pale Yellow to Yellow
  • Corrections for temperature & humidity: Temperature correction is necessary
  • Relative Standard Deviation: 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
  • Reaction Principle:
    • C6H10O + 2,4-Dinitrophenylhydrazine → Dinitrophenylhydrazone
  • 2 Year Shelf Life
  • Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.

Possible coexisting substances and their interferences:

Substance Interference Changes color by itself to
Ketones + Yellow

Other substance measurable with this detector tube:

Substance: Correction Factor: No. of Pump Strokes: Measuring Range:
Diacetone alcohol by scale 2 2.5 to 100 ppm
Furfural 1.0 4 2 to 30 ppm
Isophorone 1.0 8 2 to 30 ppm

Cyclohexanone Detection Tubes (2 to 75 ppm)