Hydrogen Sulfide (H2S) Detection Tubes (0.1 to 4.0 ppm)
Gastec Detection Tube No. 4LT measures Hydrogen Sulfide with a Range of 0.05 to 4.0 ppm.
- Measuring Range:
- 0.05 to 0.1 ppm
- (0.1) to 2.0 ppm
- 2.0 to 4.0 ppm
- Number of Pump Strokes: 2, 1, 1/2
- Correction Factor: 1/2, 1, 2
- Sampling Time: 1.5 minute per pump strokes
- Detecting Limit: 0.01 ppm (2 pump strokes)
- Changes from Yellow to Pink
- Corrections for temperature & humidity: Unnecessary
- 2 Year Shelf Life (in the refrigerator)
- Reaction Principle: Hydrogen sulfide reacts with the reagent to form intermediate material which stains indicator pink
- Relative Standard Deviation: 5% (for 0.1 to 2 ppm)
- Store the tubes in the refrigerator to keep at 10�C (50�F) or below.
| Substance |
Concentration |
Interference |
Changes color by itself to |
| Ammonia |
? 2.3 ppm |
- |
No |
| Ethyl mercaptan |
? 0.5 ppm |
+ |
Pink (? 0.4 ppm) |
| Hydrogen chloride |
? 4.5 ppm |
+ |
Pink (? 4.0 ppm) |
| Hydrogen cyanide |
? 0.05 ppm |
+ |
Pink (? 0.05 ppm) |
| Nitric acid |
? 6.0 ppm |
+ |
Pink (? 5.0 ppm) |
| Sulfur dioxide |
? 10 times |
+ (Two layers) |
Pale pink (? 2.0 ppm) |
| Nitrogen dioxide |
? 5 times |
- |
Pink (? 6.5 ppm) |
| Hydrogen fluoride |
? 12.0 ppm |
+ |
Pink (? 11.0 ppm) |
Hydrogen Sulfide (H2S) Detection Tubes (0.1 to 4.0 ppm)